Carnegie Mellon University
Browse

An Ankle-Foot Prosthesis Emulator with Control of Plantarflexion and Inversion-Eversion Torques

Download (2.58 MB)
journal contribution
posted on 2015-05-01, 00:00 authored by Steven Collins, Myunghee Kim, Tianjian Chen, Tianyao Chen

Ankle inversion-eversion compliance is an important feature of conventional prosthetic feet, and control of inversion, or roll, in robotic prostheses could improve balance for people with amputation. We designed a tethered ankle-foot prosthesis with two independently-actuated toes that are coordinated to provide plantarflexion and inversion-eversion torques. This configuration allows a simple lightweight structure with a total mass of 0.72 kg. Strain gages on the toes measure torque with less than 2.7% RMS error, while compliance in the Bowden cable tether provides series elasticity. Benchtop tests demonstrated a 90% rise time of less than 33 ms and peak torques of 180 N·m in plantarflexion and ±30 N·m in inversion-eversion. The phase-limited closedloop torque bandwidth is 20 Hz with a 90 N·m amplitude chirp in plantarflexion, and 24 Hz with a 20 N·m amplitude chirp in inversion-eversion. The system has low sensitivity to toe position disturbances at frequencies of up to 18 Hz. Walking trials with five values of constant inversion-eversion torque demonstrated RMS torque tracking errors of less than 3.7% in plantarflexion and less than 5.9% in inversion-eversion. These properties make the platform suitable for haptic rendering of virtual devices in experiments with humans, which may reveal strategies for improving balance or allow controlled comparisons of conventional prosthesis features. A similar morphology may be effective for autonomous devices.

History

Publisher Statement

© 2015 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works.

Date

2015-05-01

Usage metrics

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC